Testomat LIMIT

Online Limit Value Analyzer for Total Hardness

Key features:

  • Continuous hardness monitoring every 1 – 60 minutes
  • Set alarm point between 0.89 – 445 ppm CaCO3 total hardness
  • Touchscreen for easy set up
  • Self-calibrating
  • Low water and reagent consumption
  • Low maintenance requirements

Optimize water quality testing with the Testomat LIMIT®, an automated solution for total hardness determination that uses a precise titration method. Immediate notifications via alarms are triggered when water hardness exceeds defined limits. Set alarm point between 0.89 – 445 ppm CaCO3 total hardness
 
Testomat LIMIT® seamlessly integrates with other systems, transmitting results directly to a PLC, softener, or filters for automatic corrective actions. By proactively calculating reagent availability, the Testomat LIMIT® significantly enhances operational efficiency.

Technical Data

Measuring method Titration
Measuring range 0.89 – 445 ppm CaCO3 total hardness (see reagents below)
Self-calibrating Yes
Power supply24 V DC (power supply included)
Power consumptionMax. 16 W
Output signal4-20 mA and dry signal
Degree of protectionIP 44
Ambient temperature50 – 104°F (10 – 40°C)
Dimensions (W x H x D)17 x 13 x 6 inch
Weight9 lbs
Country of originGermany

Water Supply

Operating pressure4.4 to 15.4 psi (0.3 to 1 bar) OR
14.5 to 116 psi (1 to 8 bar)
Min. water flow14 fl oz (400ml) per minute
Water inlet1/4-inch quick connect fitting
Drain1/2 tube
Water temperature50 – 104°F (10 – 40°C)
a cooler can be installed for temperatures above 104°F