Testomat PRO FE

Highly Accurate Measuring Instrument for Iron

Key features:

  • Continuous iron monitoring every 1 – 90 minutes
  • Real-time readings in display
  • Low maintenance
  • Range of 0 – 1 ppm CaCO3 total hardness
  • Integrated SD card to store readings
  • Self-calibrating
  • Low water and reagent consumption
  • Monitoring of two measuring points
  • Self clean function

The process photometer Testomat PRO® FE is an online analyzer for monitoring dissolved iron content in water from 0 to 1 ppm using photometric analysis.

Range of Application

  • Monitoring of systems for removing iron from well water
  • Controlling industrial or drinking water

Measuring Range

  • 0.0 – 0.65 ppm (resolution 0.01)
  • 0.7 – 1.0 ppm (resolution 0.1)

Technical Data

Measuring method Photometric
Self-calibrating Yes
Power supply24 VAC OR 120 VAC OR 240 VAC
Power consumptionMax. 50 W
Output signal4-20 mA and dry signal
Degree of protectionIP 44
Ambient temperature50 – 104°F (10 – 40°C)
Dimensions (W x H x D)11 x 14 x 6 inch
Weight18 lbs
Country of originGermany

Water Supply

Operating pressure4.4 to 15.4 psi (0.3 to 1 bar) OR
14.5 to 116 psi (1 to 8 bar)
Min. water flow14 fl oz (400ml) per minute
Water inlet1/4-inch quick connect fitting
Drain1/2 tube
Water temperature50 – 104°F (10 – 40°C)
a cooler can be installed for temperatures above 104°F
ReagentTotal ironQuantity in mlOrder numberSafety data sheet
FE 2005 B0 – 1.0500156251
FE 2005 A0 – 1.050015625